Data of various lenses are included in the Contact Lens Fitting display. Can data for other lenses be imported too?
Yes, data from any lens can be manually input if the parameters are known.
Yes, data from any lens can be manually input if the parameters are known.
No, lenses cannot be individually measured. Well-known soft lenses were analyzed and the information is programmed into the software. The role of the OxiMap® is to educate patient’s on the importance of using lenses with high oxygen transmissibility, showing them ...
Topographic indices for keratoconus screening are incorporated in the software, as well as Fourier Analysis. Objective inspection of the sagittal map is another tool to aid with diagnosis.
No, the Keratograph is a Placido topographer. Placido discs are projected on the tear film, the software analyzes the resulting images and measurement data of the anterior cornea is generated. A Pacido topographer can't measure the posterior cornea on its ...
No, a keratometer is included in the hardware measuring real Ks of the anterior surface.
Yes, with the 22 Placido disc 22.000 real measured data points are generated after each measurement. None of these are extrapolated measurements.